LA Range: LAPP

Bigneat manufactures and installs LAPP enclosures for sample/process protection.

Bigneat’s LAPP enclosure has been designed following years of practical experience and is an example of Bigneat’s innovation in design & technology.
LAPP enclosures provide downflow sterile ISO Class 5 Clean air to the work surface ensuring protection of the sample/process.

Access to the enclosure interior during operation can be gained through the three lower sliding windows within the lower hinged door (access 11.8″ high) and full height sliding doors allowing for access to the robotics system inside the enclosure during use, loading and maintenance.

Interlocking upper and lower main access doors are locked down with ‘T’ handles during operation.

This enclosure’s air flow is 100% pass through system, air in and air out. Product protection is a paramount function for the LAPP model with automatic fan speed control maintaining a minimum face velocity when one door is open which complies to recognised Standards worldwide.

Features:

  • Great ergonomics with access to four sides of the robotic platform.
  • Excellent all-round visibility of the enclosed system.
  • Easy touch, state-of-the-art control system – displays enclosure status and controls airflow system, hour counter.
  • Display shows status of enclosed robotic system which is mounted to enclosure exterior.
  • Audible and visual alarm indication of low airflow.
  • Door open warning alarm.
  • Display/lighting beacons show status of enclosed robotic system.

 

Options and extras

  • Additional electrical sockets to suit robotics system.
  • Computer shelf on flexible arm.
  • Hydrogen peroxide or alternative fumigation connections, night doors, removed for normal operation.
  • Under-bench storage and shelving to suit.
  • Universal control panel.
  • UV lighting, linked to timed on/off facility in control system.
  • Variety of door options.
  • Waste tips and plates chute and ventilated waste container.

Performance and Standards

Meeting recognised Standards worldwide.
ANSI/IEST/ISO 14644- 1:1999: Part 1: Para 2.4.1.

SEND ENQUIRY ABOUT THIS PRODUCT

LA Range: LAPP

Enclosure general construction:

Gloss finish epoxy-coated, steel frame modules for the enclosure and lower stand.
Epoxy coated, steel doors, fitted with neoprene seals.
Clear acrylic, 8mm panels fitted to UHMW triple tracking.
Lab grade, HPL 16mm composite laminate work surface.

Model External Dimentions (WxDxH) Internal dims (WxDxH) Inflow air velocity (fpm) Downflow air velocity per hr
RP1500/01 Nominal 59″x46.5″x94.5″
(Design 1500 x 1180 x 2400)
Nominal 56″x43.5″x45″
(Design 1420 x 1100 x 1144)
>79 >60
RP1800/01 Nominal 71″x46.5″x94.5″
(Design 1800 x 1180 x 2400)
Nominal 68″x43.5″x45″(Design 1720 x 1100 x 1144) >79 >60
RP2200/01 Nominal 86.5″x46.5″x94.5″
(Design 2200 x 1180 x 2400)
Nominal 83.5″x43.5″x45″(Design 2120 x 1100 x 1144) >79 >60
RP2600/01 Nominal 102″x46.5″x94.5″
(Design 2600 x 1180 x 2400)
Nominal 99″x43.5″x45″(Design 2520 x 1100 x 1144) >79 >60
RP3200/01 Nominal 126″x46.5″94.5″
(Design 3200 x 1180 x 2400)
Nominal 123″x43.5″x45″(Design 3120 x 1100 x 1144 >79 >60

SEND ENQUIRY ABOUT THIS PRODUCT

LA Range: LAPP

Pre-filtration eliminates particles at 5.0μm or larger to an efficiency of 92% as defined in BS EN ISO 779.

Particulate filtration HEPA filtration (H14 Standard) eliminates particles 0.3 μm or larger to an efficiency of 99.995%.


Free Risk Assessment Service
What are your risks? Please contact us for a free assessment and our chemist will undertake a risk analysis and suggest the best filter for your purposes. Please click on the download button to the right.

SEND ENQUIRY ABOUT THIS PRODUCT